DMS500

The DMS500 has established itself since 2002 as the particle measurement instrument of choice for engine research, development and calibration, with hundreds of customers worldwide including OEMs, Tier 1s and Universities. Successive generations of the DMS500 have offered improved capabilities and made the DMS500 the most dependable instrument in its class.

Flexible measurements – fit for future development

Heavy Duty Diesel size distributionThe DMS500 represents a "turn-key" system. The fully integrated dilution and sample conditioning system allows measurment from from the exhaust port, pre- or post- after-treatment, in the CVS tunnel and even from ambient.

10Hz data and a T10-90% response time of just 300ms even with a 7 metre sample line allow the DMS500 to offers valuable insights into engine operation, even during highly transient cycles such as RDE, NRTC and FTP75.

Particle Mass (PM) for gravimetric regulations and for engine out model development (e.g. DPF/GPF load models).

Particle Number (PN) with a fully characterised thermodilution system including a Catalytic Stripper to remove volatile particles when required. Flexible software allows measurements with a 23nm roll off (Euro 6) or the possible future 10nm roll off (possible Euro 7).

Particle Size distributions (5nm -2500nm) for maximum understanding of combustion behaviour, particle formation and evolution, and the effects of engine controls on PN and PM.

Developed for daily use – reliable results

Backed by over 15 years of continuous development, the DMS500 is straightforward to operate even for inexperienced users.

Dependable error detection and optional test cell integration allow the DMS500 to be seamlessly added as part of general test cell equipment.

Traceable calibration certificates and automated checks guarantee accurate and repeatable results.

Learn more:

Download a brochure and contact Cambustion to discuss your application.

Key Application Notes: 
Title Data type Download File Size Last updated
Real-time aerosol size, number and mass measurements from an electronic cigarette Application note PDF icon dms12v01.pdf 120.11 KB 24-Apr-2014
Real-time particle size measurements of heat-not-burn tobacco device Application note PDF icon dms15v01.pdf 870.75 KB 09-Oct-2014
DMS500 Installation Guide for Engine Sampling Application note PDF icon dms11v02.pdf 512.72 KB 02-Feb-2015
Gasoline Direct Injection Particulate Measurement with the DMS Series Application note PDF icon dms08v01.pdf 919.53 KB 24-Feb-2015
Flame Sampling with DMS500 Application note PDF icon dms07v01.pdf 260.74 KB 24-Feb-2015
Real-time mode finding and lognormal fitting with DMS series fast particulate spectrometers Application note PDF icon dms06v03.pdf 682.83 KB 24-Feb-2015
Measurement of DPF/CRT performance with DMS500 Application note PDF icon dms05v02.pdf 1.2 MB 24-Feb-2015
DMS500 measurement of drug delivery inhalers Application note PDF icon dms04v02.pdf 85.64 KB 24-Feb-2015
DMS500 sampling options Application note PDF icon dms03v04.pdf 511.06 KB 24-Feb-2015
Measurement of Volatile Aerosols with the dms500 Application note PDF icon dms02v01.pdf 45.93 KB 24-Feb-2015
Ambient aerosol sampling with DMS500 Application note PDF icon dms10v01.pdf 295.83 KB 24-Feb-2015
Particle number measurement with the DMS Series Application note PDF icon dms09v01.pdf 351.19 KB 24-Feb-2015
Brake particle measurements Application note PDF icon dms17v01.pdf 529.01 KB 08-Mar-2016
Real-time particle size measurements of user triggered electronic cigarette smoke Application note PDF icon dms14v02.pdf 910.84 KB 16-Aug-2016
Nanoparticle emissions from cooking Application note PDF icon dms19v01 Nanoparticle emissions from cooking.pdf 1.94 MB 10-Oct-2016
Water pipe smoking Application note PDF icon dms22 Water pipe smoking.pdf 931.87 KB 04-Aug-2017
Number : Size Spectrum Correlation with Particulate Mass Measurements Application note PDF icon dms01.pdf 890.73 KB 13-Nov-2017