Fast FID Principles

The flame ionisation detector (FID) is the industry standard method of measuring hydrocarbon (HC) concentration.

 

The sample gas is introduced into a hydrogen flame inside the FID. Any hydrocarbons in the sample will produce ions when they are burnt. Ions are detected using a metal collector which is biased with a high DC voltage. The current across this collector is thus proportional to the rate of ionisation which in turn depends upon the concentration of HC in the sample gas.

The ionisation process is very rapid, so the slow time response of conventional FIDs is mainly due to sample handling. A typical slow analyser might have a response time of 1-2 seconds.

The Cambustion HFR fast response FID analyzers use conventional detection principles and a unique patented sampling system to give millisecond response times.

The Cambustion fast FID consists of a main control unit (MCU) and two remote sampling heads (which house the FIDs). The dual channel nature of the instrument enables simultaneous real-time measurement in two locations allowing, for example, evaluation of catalyst performance.

Contact Cambustion for more information and prices.